+62 (031) 8284914, 085100605225, 085100974976, Fax. +62 (031) 8284955 sumberarumabadi@gmail.com

C5010 & C5020 Multi parameter analysers MEREK CONSORT BELGIA

C5010 C5020 C5010 package

pH
Multi-point (1…3) calibration with up to three buffers out of eleven
pre-program med pH buffers.
Reads pH with 0.01 pH resolution.
mV
Features mV calibration for accurate ORP mea su rements.
Reads potentials with 1 mV resolution.
Conductivity
Use a 1 cm
-1
electrode to measure from 0.1 µS/cm to 100 mS/cm.
Use a 0.1 cm
-1
electrode to measure from 0.01 µS/cm to 10 mS/cm.
(C5020 only)
Use a 10 cm
-1
electrode to measure from 1 µS/cm to 1000 mS/cm.
(C5020 only)
Automatically se lects cor rect rangeand frequency.
Selectable reference temperature: 20° or 25°C.
One-point calibration.
Dissolved oxygen
Operates with a galvanic oxygen electrode requiring no polarisation
time and no zero calibration.
Reads dissolved oxygen with 0.01 mg/l or 0.1% resolution.
Rapid air calibration.
Temperature
Reads temperatures with 0.1°C resolution.
Manual or automatic temperature compensation.
Calibrates temperature probe for quality measurements.
Inputs
One common input for pH and mV.
One common input for conductivity and dissolved oxygen.
One input for a Pt1000 automatic temperature probe.
Display
Bright LCD screenfor better readability.
A  white  backlight  automatically  illuminates  when  operated  on  the
mains.
Stability  indicator  prompts  the  user  when  readings  should  be
taken.
The interactive LCD screen provides step by step instructions in the
language of your choice (English, Dutch, French, German).
Shows a GLP report on the LCD screen.
Data-logging(C5020 only)
Storage memory for 300 values including temperature.
Cabinet
Robust dust and splash-proof cabinet.
Special features
Three year warranty.
Mains and rechargeable battery ope ration with programmable automatic switch-off.
Optional 12 V car adaptor.
Pre-programmed standards
pH  buffers: 1.68,  2.00,  4.00,  4.01,  6.87,  7.00,  9.18,  9.21,  10.01,
12.00, 12.45 (at 25°C).
Conductivity: 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm (at 25°C).
One pH/mV channel
One conductivity/oxygen channel
One temperature channel
C5010 •C5020 Multi-parameter analysers

One pH/mV channel
One mV-H2 One mV-H2 One mV-H/mV channel
One temperature channel
rH
2 rH
2 rH
Bio-elec tronic multimeter for the study of
the biological water quality according to
Vincent’s method.
mV-H2 mV-H2 mV-H
Features mV calibration for accurate ORP
mea su rements.
Reads potentials referred to the stan dard
hydrogen electrode with 1 mV resolution.
Connect a special pH/ORP combination electrode to the meter and reads rH
2
directly on
the screen.